Gloved Laminar Air Flow(GLAF)
Protection: Product Operator Environment
Conditions of use
The GLAF is a real alternative to the use of isolators, i.e. it enables the containment of sterile but non toxic operations, by minimizing human intervention, which may contaminate the products.
Working principle
A class ISO 5 (NF, EN, ISO 14644-1) laminar air flow is blown vertically from the plenum towards the ground, and is guided within the area to be protected via a system of rigid transparent polycarbonate panels. This air flow maintains the area to be protected under a greater pressure than the surrounding room,
Configuration
Selection of accessories to complete the modules being assembled according to the clean area
configuration. Includes:
Automatic regulation
Light
Manometer
PVC gloved containment curtains
Totally constructed in stainless steel 304 with diffusion and protection grid for pre-filter and HEPA filter or powder coated steel.
|
|
手套层流罩
保护对象: 产品 操作人员 环境
使用场合
层流手套筒是隔离器的经济替代设备。减少由于人员操作导致对无菌(无毒)产品的污染。
工作原理
通过从上至下的百级垂直层流和设备外周带手套的局部围档,形成相对正压的保护
配置 以下附件为标准配置:
调节器(风速)
照明灯具
压差表
PVC软帘手套箱
框架可以是整体304不锈钢或钢板静电喷涂。吸风口有初效过滤器保护。
|